Product Guides2026-5- 22

Key Points of This Issue:

TNO, together with the Dutch National Metrology Institute (VSL), used an RGM (Reference Gas Mixture) system to conduct systematic laboratory calibration of the HT8700E. A custom optical calibration cell was used to temporarily enclose the open optical path, and validation was completed over a concentration range of 0–1262 ppb and a relative humidity (RH) range of 7–55%. Core findings: in the <200 ppb concentration range (i.e., the typical concentration range downwind of farms), the HT8700E showed excellent linear response (R² = 0.999), and the humidity effect was not significant; at ≥200 ppb, humidity effects appeared, and an RH correction term needed to be introduced. Based on this, TNO established a zone-specific calibration strategy and a second-order temperature compensation model, laying a metrological foundation for field applications.


Composition of the Calibration System

Open-path instruments cannot be directly connected to standard gases, which is a core bottleneck in their value traceability. To solve this problem, TNO collaborated with VSL to design a closed-path calibration scheme specifically for open-path instruments.

VSL’s RGM system uses two-stage thermal mass flow controllers (MFCs) for dynamic dilution of high-pressure standard gas, with a maximum dilution factor of about 800 times. It can output ppb-level NH ₃ to the calibration cell and independently regulate the RH of the output gas. The relative expanded uncertainty (coverage factor k = 2) of the output NH ₃ concentration is about 2.7%.

In response to the open-path characteristics of the HT8700E, TNO designed a custom optical calibration cell (volume 3.7 L, surface area 1500 cm²) to temporarily enclose the open optical path between the HT8700E laser path and the mirror. RGM output was injected from one end of the calibration cell and exhausted from the other end by a constant-flow pump. The flow rate was maintained at 3.5 L/min, and the theoretical residence time of the sample was about 1 minute.

image.png


Experimental Design

TNO calibrated the HT8700E installed on a mobile monitoring vehicle platform under laboratory conditions. The concentration settings covered the full range from background to high concentration: 0, 29, 72, 143, 285, 710, and 1262 ppb; humidity covered four RH plateaus: 7%, 25%, 47%, and 50%. The highest concentration point (1262 ppb) was scanned in both upward and downward concentration directions under 50–55% RH conditions to evaluate adsorption hysteresis effects.

To ensure data reliability, the study adopted a two-level data quality control strategy:

  • Simple filtering: data stability (degree of fluctuation) was evaluated using a 30-minute moving window, requiring RGM ≤ 1%, RH ≤ 3%, and instrument reading ≤ 3%;

  • Strict filtering: after the start of each stable stage, the first 15 minutes of transition data were removed, and the subsequent data were detrended. Only the last 1 hour was retained for analysis;

Statistical test results showed that in the low-concentration range, there was no significant difference between the results obtained by the two screening methods (t-test, p ≥ 0.05), indicating that the system response was stable.


Core Validation Results in the Low-Concentration Range (NH ₃ < 200 ppb)

Within the typical concentration range for mobile measurements downwind of farms (i.e., <200 ppb), the HT8700E demonstrated excellent metrological performance:

  • Calibration equation: reference concentration = (raw reading − 0.16) / 0.505

  • Coefficient of determination: R² = 0.999

  • Standard error of slope: 0.002

  • Standard error of intercept: 0.12 ppb

This result indicates that within this concentration range, the HT8700E response is highly linear, and data points at different humidity levels are distributed on the same regression line. The humidity effect is not statistically significant. This significantly simplifies the field data processing workflow—because in actual mobile measurements, the HT8700E samples directly in open-path mode, and its insensitivity to humidity changes in the low-concentration range usually means that complex humidity correction is not required during data processing.


Humidity Effects in the High-Concentration Range (VSL ≥ 200 ppb)

When the concentration increased to ≥200 ppb, under enclosed calibration cell conditions, NH ₃ concentration showed a significant RH dependence. The physical mechanism is that as humidity increases, the surface water film gradually thickens, and the dissolution and adsorption processes of NH ₃ in the water film are enhanced, causing measured values to be systematically lower than the input concentration; at the same time, hysteresis effects occur during the concentration ramp-up and ramp-down stages.

Quantitative data showed that at an input concentration of 285 ppb, when RH increased from 8% to 50%, the measured value decreased from 152 ppb to 143 ppb; at 1267 ppb, the measured value showed a significant negative correlation with RH (R² = 0.988). Based on the above characteristics, TNO established a calibration model considering humidity effects and introduced an interaction term between concentration and RH (i.e., their coupling term) into the regression equation, with an overall goodness of fit of R² = 0.999.

It should be particularly emphasized that TNO clearly pointed out that the above high-concentration RH effect was observed only under custom optical calibration cell conditions and should not be directly applied to field open-path data.

image.png

image.png


Discovery and Modeling of Temperature Drift

TNO conducted a 48-hour multi-concentration calibration of the HT8700E outdoors and carried out a 24-hour temperature stability test at a fixed concentration of 50 ppb. The results showed that ambient temperature changes affect the laser cavity temperature, introducing concentration drift. Further analysis found a hysteresis loop effect between the internal instrument temperature and the cooling water circulation temperature (the response paths of the two were inconsistent during heating and cooling), indicating that the system temperature control response has a certain lag, which is also one of the important causes of drift (this hysteresis is caused by adsorption in the gas cell tubing under calibration conditions and is fundamentally different from the open-path state under actual test conditions).

TNO established a second-order polynomial compensation model based on the HT8700E’s built-in water circulation temperature sensor (Tw). After correction, the slope was 0.507 and the intercept was 0.18 ppb, almost consistent with the laboratory calibration results (slope 0.505, intercept 0.16 ppb). This correction model relies only on the HT8700E instrument’s own temperature sensor and does not require additional meteorological equipment, providing strong feasibility for field operation.


Conclusion:This series is based on the TNO technical report “Traceable Calibration of Open- and Closed-Path Ammonia Analyzers Using the VSL RGM System and Its Impact on Livestock Emission Uncertainty” (TNO 2024 R12917), which is an official deliverable of the EURAMET quantiAGREMI project (21GRD10). Figure and table numbers retain their original labels to facilitate readers’ retrieval of the original literature. For the complete report PDF or further technical exchange, please contactinfo@healthyphoton.com.

Contact Healthyphoton
  • (+86) 400-961-6990

  • info@healthyphoton.com

  • Room 305, Building 1, Zhongchuang Science Park, Jinyuan Road, Panhuo Street, Yinzhou District, Ningbo, China

Full Name*
Phone Number*
Email Address*
Please Enter Full Name!
Please Enter A Valid Mobile Phone Number!
Please Enter A Valid Email Address!
Submission Successful
Submission Failed

Speak with our experts to find the right sensing solution for your application.

Innovation for Green Life

Precision sensing solutions for environmental monitoring, scientific research, and industrial applications.

Scan The Code To Follow Us

Scan The Code To Follow Us

? HealthyPhoton Technology Co., Ltd. All Rights Reserved.Privacy PolicyTechnical Support:Chemical Instrument Networksitemap.xml

(+86) 400-961-6990

info@healthyphoton.com

Scan the code to add WeChat