Resources
HomeResourcesProduct GuidesHT8700E Breaks Through the Technical Bottlenecks of High-Precision Ammonia Monitoring
Key Points in This Issue
Ammonia (NH₃) is a core precursor of nitrogen deposition and atmospheric secondary pollution. Globally, livestock and poultry farming is considered the largest anthropogenic emission source of NH₃. However, the “stickiness” of NH₃, its humidity dependence, and its low concentration in the background atmosphere have long made accurate measurement face metrological traceability challenges. The European Association of National Metrology Institutes (EURAMET) Joint Research Project quantiAGREMI (project No. 21GRD10) aims to improve the metrological traceability and cross-platform consistency of NH₃ measurements in livestock farming. Healthy Photon’s HT8700E Open-Path NH₃ Analyzer (HT8700E is the EU version of HT8700) was selected by TNO in the Netherlands as one of the key NH₃ sensors in its mobile remote emission measurement system, playing a key role in moving from “technically available” to “metrologically traceable.”
Importance and Difficulties of Ammonia Monitoring
Ammonia (NH₃) plays a central regulatory role in atmospheric chemistry: it is both an important contributor to nitrogen deposition and a key precursor for the formation of secondary inorganic aerosols such as ammonium sulfate and ammonium nitrate. According to estimates by the United Nations Environment Programme, more than 80% of global NH₃ emissions come from agricultural activities, among which livestock and poultry farming (animal housing, manure storage, and application) is absolutely dominant. Accurately quantifying farm-scale NH₃ emissions has irreplaceable scientific value for assessing the health impacts of air pollution, supporting integrated assessments of greenhouse gases and the nitrogen cycle, and verifying the effectiveness of emission reduction measures.

However, accurate NH₃ measurement has long faced three technical challenges.
First, surface adsorption effects. NH₃ molecules have significant polarity and strong surface interaction capability, and are prone to adsorb onto the inner walls of instrument tubing, optical windows, and cavity surfaces, causing significant sampling delay and surface memory effects. Although traditional closed-path analyzers have excellent precision, relatively long sampling lines may still introduce significant response lag and signal tailing. This problem is particularly prominent in rapid dynamic measurements on mobile platforms.
Second, humidity cross-sensitivity. NH₃ is highly sensitive to water vapor conditions. Changes in ambient humidity not only affect the absorption spectrum but also change spectral line shapes through pressure broadening effects. In the case of a closed optical cavity, high humidity may also form a water film on the optical cavity surface, enhancing the adsorption/desorption processes of NH₃ and thus leading to response lag and additional measurement uncertainty.
Third, low concentration and lack of traceability. NH₃ concentrations downwind of farms are often at the ppb level (background ~2 ppb, plume peak ~50 ppb), close to the limit of detection (LOD) of some instruments. Traditional approaches make it difficult to accurately obtain ppb-level standard gases, and there is still a lack of a unified and widely recognized value traceability method, which limits the comparability of data between different instruments and different laboratories.
It is precisely against this background that EURAMET launched the quantiAGREMI project (project No. 21GRD10) in 2021, aiming to establish a metrological traceability system from reference gas standards and sensor calibration to emission inversion. TNO, as one of the important participating institutions of the project, participated in the development and validation of mobile remote measurement technology.
Technical Positioning of the HT8700E
In its mobile measurement platform, TNO selected Healthy Photon’s HT8700E Open-Path NH₃ Analyzer as one of the core NH₃ sensors. This choice was based on the following technical features of the HT8700E:
Open-path design: sampled air does not need to pass through any inlet tubing, significantly reducing adsorption and response lag caused by tubing, and achieving real-time sampling at up to 10 Hz and a concentration detection limit of 0.3 ppb;
Mid-infrared quantum cascade laser: selects the core region of a strong NH₃ absorption band and reduces spectral interference with major atmospheric constituents such as H₂O through line selection, ensuring high selectivity and sensitivity;
Adaptability to mobile platforms: its size and power consumption are suitable for vehicle-mounted or trailer platforms and can be installed on a vehicle roof, directly exposed to ambient air, for gas plume tracking and spatial distribution measurements.

In TNO’s mobile monitoring vehicle system, the HT8700E was installed on the roof of the trailer (with a dedicated rainproof metal cover), complementing the closed-path laser analyzer inside the vehicle—the QCL-TILDAS dual-laser analyzer (responsible for CH₄ and N₂O measurements).

The mobile monitoring vehicle traveled back and forth along roads downwind of the farm, crossing the emission plume—that is, the air mass formed by emitted gases under the action of wind fields and atmospheric turbulence with concentrations higher than background. Combined with N₂O tracer release, the overall NH₃ and CH₄ emission rates of the farm were estimated through the Tracer Dispersion Method (TDM). During the rapid crossing of the plume cross-section by the mobile platform, the 10 Hz high-frequency sampling of the HT8700E can finely capture instantaneous NH₃ concentrations, while the open-path design significantly reduces signal tailing and peak shift caused by sampling adsorption in closed-path systems, improving the quantitative accuracy based on the plume integration method.
Conclusion: This series is based on the TNO technical report “Traceable Calibration of Open- and Closed-Path Ammonia Analyzers Using the VSL RGM System and Its Impact on Livestock Emission Uncertainty” (TNO 2024 R12917), which is a formal deliverable of the EURAMET quantiAGREMI project (21GRD10). Figure and table numbering retain the original labels to facilitate readers’ retrieval of the original literature. For the complete report PDF or further technical exchange, please contact info@healthyphoton.com.
(+86) 400-961-6990
info@healthyphoton.com
Room 305, Building 1, Zhongchuang Science Park, Jinyuan Road, Panhuo Street, Yinzhou District, Ningbo, China
Speak with our experts to find the right sensing solution for your application.
Innovation for Green Life
Precision sensing solutions for environmental monitoring, scientific research, and industrial applications.? HealthyPhoton Technology Co., Ltd. All Rights Reserved.Privacy PolicyTechnical Support:Chemical Instrument Networksitemap.xml