Key Points of This Issue:
During the quantiAGREMI joint observation campaign, TNO conducted in-situ field calibration of the HT8700E at a dairy farm in the Netherlands using VSL’s RGM system, and simultaneously calibrated the UK Centre for Ecology & Hydrology (UKCEH) HT8700E and two Aerodyne closed-path analyzers. The results showed that, after temperature correction, the field calibration slope of the TNO HT8700E (0.507) deviated by less than 0.4% from the laboratory value (0.505); the HT8700E showed a similar linear response, and its signal noise level was within the typical range for instruments of this type. Decomposition of the uncertainty in gas plume signals showed that the relative uncertainty of HT8700E plume measurements was approximately 5.5%—of which signal noise contributed 88%, and calibration factors contributed less than 6%. This indicates that, after calibration, data quality is mainly limited by the instrument’s own background noise level, while the error introduced by the calibration process itself is extremely low. Overall measurement accuracy meets the routine requirements for field plume monitoring (Note: a relative uncertainty of 5.5% is acceptable for open-path plume measurement scenarios with significant environmental disturbances).
Experimental Setup
Laboratory calibration established a traceable calibration basis for the HT8700E, but the instrument’s stability in real field environments and the comparability among different types of instruments are the key factors determining its scientific application value. Against this background, TNO conducted a multi-instrument synchronous calibration and comparison experiment during the joint observation campaign at a dairy farm in the Netherlands.

TNO’s calibration of the HT8700E was implemented in three stages: 1. Farm pre-calibration: at the farm site, the RGM system was used for dry-state (RH<5%) calibration at three concentration points: 0, 50, and 200 ppb. Due to strong winds, the sealing at both ends of the calibration chamber may have been disturbed, and data quality declined somewhat; therefore, this result was mainly used for rapid on-site checking. The calibration slope was 0.428, and the intercept was −2.20 ppb. 2. Petten post-calibration: after returning to the TNO site, a wider range of 0, 20, 50, 100, 200, 496, and 750 ppb was used for repeated verification under dry and semi-dry (RH≈33%) conditions. Under dry-state calibration, the slopes were 0.492/0.507 (before/after correction), and the intercepts were −2.35/0.18 ppb (before/after correction). 3. 24-hour temperature stability test: temperature drift was tracked at a fixed concentration of 50 ppb, supporting the second-order temperature compensation model described.

The dry-state intercepts of the pre-calibration and post-calibration were −2.20 and −2.35 ppb, respectively, which were highly consistent; the temperature-corrected slope of the post-calibration was 0.507, with a relative deviation of less than 0.4% from the laboratory calibration value of 0.505. This result strongly demonstrates that the metrological performance of the HT8700E maintained excellent stability throughout the entire process from the laboratory to the farm and back, and that temperature change was the main additional compensation factor.
Side-by-Side Comparison of Co-Located Instruments
TNO used the VSL RGM system to synchronously calibrate the following instruments:

*TI-QCL: Johann Heinrich von Thünen Institute, Germany, established in 2008. It is one of four federal research institutes under the German Federal Ministry of Food and Agriculture (BMEL).
Key Finding: Consistent Response Pattern of HT8700E Series Products
Although the three HT8700E units (TNO mobile, TNO flux, and UKCEH) had different slopes (reflecting different individual response coefficients of the devices), all showed a linear response highly consistent with the VSL standard (R²≥0.999). This means that, as long as in-situ calibration is performed using the RGM or an equivalent traceable standard, data from HT8700E series devices can achieve a unified metrological basis and cross-platform comparability. Differences in original slopes (in the range of 0.3–1.0) can be fully corrected through a single multi-concentration calibration.
The UKCEH HT8700E showed some noise fluctuations and stability differences in this test (Note: this difference is still within the allowable error range for typical environmental monitoring applications and does not affect data validity or trend judgment), which may be related to the unit’s operating hours, maintenance status, or on-site temperature conditions. Under semi-dry conditions (20–30% RH), its calibration slope decreased from 0.454 under dry conditions to 0.337. This phenomenon is very likely related to enhanced NH₃ adsorption/dissolution processes after the water film on the surface of the custom optical calibration chamber increased. This trend is consistent with response changes previously observed for the TNO HT8700E.
Uncertainty Decomposition of Plumes
The TNO HT8700E obtained 21 groups of NH₃ plume observation samples on a road northwest of a dairy farm (about 150 m from the barns). Under meteorological conditions of northeasterly wind (50.1°±6.3°) and wind speed of 6.9±0.9 m/s, NH₃ emitted from the barns was advected by the mean wind field and affected by turbulent diffusion in the boundary layer, forming distinguishable concentration enhancement signals downwind. The HT8700E recorded the concentration time series during each traverse at a 10 Hz sampling frequency. The plumes showed a typical “rise—peak—decline” structure along the vehicle-mounted sampling trajectory, with a duration of about 20–30 seconds. A systematic uncertainty analysis was conducted on the 21 plume samples, and the following statistical results were obtained:
Metric | Mean ± SD |
Net plume integral | 255 ± 15 ppb |
Absolute error | 14.3 ± 0.6 ppb |
Relative error | 5.5 ± 0.3% |
Item-by-item decomposition of the error sources for one plume signal revealed the structural characteristics of the HT8700E measurement uncertainty:
Raw signal noise: 88% (absolutely dominant, reflecting the instrument background noise under 10 Hz high-frequency sampling)
Baseline uncertainty: 7%
VSL calibration uncertainty + temperature-correction polynomial RMSE + slope error + intercept error: <6%
This error structure conveys important information: after traceable calibration with the VSL RGM and implementation of temperature compensation, the quality of HT8700E NH₃ data is mainly determined by the instrument’s own signal noise level, and the total contribution of the calibration chain (slope, intercept, VSL calibration) is less than 6%. In other words, the error introduced by the calibration system itself has been compressed to an extremely low level. The direction for further improving accuracy should focus on improving the signal-to-noise ratio (such as extending integration time or optimizing spectral fitting algorithms), rather than redesigning the calibration workflow.

Conclusion: This series is based on the TNO technical report “Traceable Calibration of Open- and Closed-Path Ammonia Analyzers Using the VSL RGM System and Its Impact on Livestock Emission Uncertainty” (TNO 2024 R12917), which is an official deliverable of the EURAMET quantiAGREMI project (21GRD10). Figure and table numbers retain their original labels to facilitate readers in retrieving the original literature. For the complete report PDF or further technical exchange, please contact info@healthyphoton.com.