Key points of this issue:
TNO used VSL-traceably calibrated HT8700E data to remotely quantify NH ₃ emissions from a Dutch dairy farm via the tracer diffusion method (TDM). Based on a weighted average of 21 plume samples, NH ₃ emissions of 0.069 ± 0.001 g/s and CH ₄ emissions of 0.623 ± 0.009 g/s were obtained, with an NH ₃:CH₄ emission ratio of 0.110 ± 0.004. This result deviates by <10% from independent indoor measurements by Wageningen University & Research (WUR) during the same period (NH ₃: 0.075 g/s), verifying the scientific reliability of the calibrated HT8700E + TDM methodological system. This marks that open-path NH ₃ analyzers have moved from “technically usable” to “metrologically traceable.”
After laboratory traceable calibration and field in-situ calibration tests, the application performance of the HT8700E was systematically evaluated. The next key question was: how can these high-precision concentration data be converted into credible farm-scale emission fluxes? TNO demonstrated its feasibility through the tracer diffusion method (TDM).
Principles of the TDM Method
The core assumption of the tracer diffusion method is the similarity of atmospheric turbulent diffusion: if a relatively stable and identifiable tracer gas (N ₂ O was used in this study) is released near an emission source at a known rate, the tracer gas and the target gases (NH ₃, CH₄) will undergo similar diffusion, dilution, and transport processes downwind. Therefore, by comparing the integrated plume concentrations of the two and combining this with the known tracer release rate, the target source emission flux can be inverted:
Q_source = Q_tracer × (M_source / M_tracer) × (∫S_source / ∫S_tracer)
where Q is the mass emission rate (g/s), M is the molar mass (g/mol), and ∫S is the time integral of the concentration enhancement (ppb·s) measured by the mobile monitoring vehicle when crossing the plume. The physical meaning of the plume integral is to accumulate the total amount of target gas passing through a unit-width plume cross-section per unit time, thereby representing the emission strength of the source at that moment.
On October 2, 2024, TNO conducted a TDM experiment at a dairy farm in the Netherlands. Meteorological conditions that day were: wind direction 50.1° ± 6.3° (northeasterly), wind speed 6.9 ± 0.9 m/s, relative humidity 71.9% ± 1.5%, and temperature 14.3–15.2°C—overall meteorological conditions favorable for plume transport and near-surface mixing. Two N ₂ O tracer release points were deployed near the barn source area, with a release rate of 0.1 g/s. The mobile monitoring vehicle traveled back and forth along the road northwest of the farm, crossing the downwind NH ₃ plume, and obtained a total of 21 valid plumes, covering two tracer release periods.

On October 2, due to the northwesterly wind, plume measurements were conducted along the road on the right side of the farm (indicated by the purple line, with purple dots showing the N ₂ O tracer placement locations); on October 11, due to the southwesterly wind, measurements were conducted on the left side of the road (indicated by the orange line, with orange dots showing the N ₂ O tracer placement locations).
Data Processing Workflow
The data processing for each NH ₃ plume included the following key steps:
Baseline correction: Baseline definition is critical to the plume integration result because its position directly affects the net plume area. TNO reconstructed the baseline using 60 seconds of background data before and after each plume.
VSL calibration: The low-concentration calibration equation obtained from laboratory calibration was applied to convert the raw signal into NH ₃ concentration.
Temperature compensation: A second-order polynomial model of the water circulation temperature Tw was introduced to compensate for ambient temperature drift.
Net plume integration: The calibrated concentration was integrated over time above the reconstructed baseline. This integral represents the total accumulated NH ₃ concentration enhancement during the measurement vehicle’s cross-section of a single plume and is the core input for TDM emission calculations.
Uncertainty propagation: The contributions of six error sources—signal noise, slope, intercept, RMSE of the temperature correction polynomial, VSL calibration, and baseline definition—were calculated item by item and combined into the total uncertainty. Because the signal characteristics of each plume differ, their integration uncertainties also differ; therefore, TNO calculated the error independently for each plume signal before performing weighted averaging.
Emission Estimation from Plume Data
After calculating the NH ₃ and CH ₄ emission rates for each plume signal separately, a weighted averaging strategy was adopted (using the inverse of each plume’s uncertainty as the weight), which can more effectively utilize information from high signal-to-noise ratio plumes than simple averaging:

Based on the 21 plume measurement data collected on October 2, 2024, the total uncertainty of the dairy farm ammonia (NH ₃) and methane (CH ₄) emissions determined by the tracer diffusion method was estimated.

Cross-Validation with Independent Methods
Indoor barn measurements conducted by Wageningen University & Research (WUR) showed NH ₃ emissions of 0.075 g/s and CH ₄ emissions of 0.68 g/s. The TNO remote measurement results were compared with these:
NH₃ deviation: −8% (0.069 vs 0.075 g/s)
CH₄ deviation: −8.4% (0.623 vs 0.68 g/s)
Emission ratio consistency: 0.111 (TNO) vs ~0.110 (WUR)
Under different institutions (TNO vs WUR) and different instruments (HT8700E vs conventional NH ₃ analyzers), the NH ₃ emission quantification results deviated by <10%, and the emission ratios were almost completely consistent. This cross-validation strongly demonstrates that VSL-traceably calibrated HT8700E data, combined with the TDM methodological framework, can provide farm NH ₃ emission estimates that are comparable to standard indoor methods but less intrusive.
Scientific Significance and Implications
For the precise verification of NH ₃ emissions from livestock and poultry farming, the application of the HT8700E gives regulators and research teams a new technical option: without entering the farm, only 1–2 hours of mobile plume transect crossings on surrounding roads are needed to obtain plume signals of gas concentrations and thereby estimate daily farm emissions. Moreover, the open-path design efficiently avoids the plume signal tailing and peak displacement problems caused by inlet adsorption in closed-path systems, making it suitable for high-humidity, rapidly changing field environments.
For users, TNO’s research results also provide a directly applicable operational paradigm: establish calibration relationships using traceable standards and simultaneously assess the impact of the environment on detection capability; in mobile plume measurements, attention should be paid to baseline reconstruction and uncertainty evaluation; at the result output stage, the robustness and credibility of emission estimation results can be improved through repeated multi-plume observations and weighted statistics. These practical experiences provide a clear path for the HT8700E to move from a “concentration monitoring tool” to a “quantitative emission analysis platform.”
Conclusion:This series is based on the TNO technical report “Traceable Calibration of Open- and Closed-Path Ammonia Analyzers Using the VSL RGM System and Its Impact on Livestock Emission Uncertainty” (TNO 2024 R12917), which is a formal deliverable of the EURAMET quantiAGREMI project (21GRD10). Figure and table numbering are retained as in the original for readers to retrieve the original literature. For the complete report PDF or further technical exchange, pleasecontact info@healthyphoton.com.